High-performance liquid chromatography as a tool to evaluate the performance of the catalytic wet peroxide oxidation of 4-nitrophenol: pre-validation of analytical methods

dc.contributor.authorRibeiro, Rui
dc.contributor.authorSilva, Adrián
dc.contributor.authorGomes, Helder
dc.contributor.authorFaria, Joaquim
dc.date.accessioned2015-11-17T11:04:59Z
dc.date.available2015-11-17T11:04:59Z
dc.date.issued2015
dc.description.abstractNitrophenol (4-NP) in aqueous solutions with concentration in the range 0.0495-118.8 mg L-1 was developed and validated under the typical criteria for in-house pre-validations. Accordingly, linearity was demonstrated through the Fisher’s exact probability test. Accuracy and precision were then assessed in three concentration levels over the linear range. The reproducibility of the catalytic wet peroxide oxidation (CWPO) of 4-NP was evaluated together with the analytical error of 4-NP determination in independent experiments. In this case, the sum of both contributions to error never reached 3%. Two multi-component HPLC methods were also developed for the determination of possible aromatic intermediates (hydroquinone, 1,4-benzoquinone, catechol, 4-nitrocatechol and phenol) and carboxylic acids (oxalic, formic, malic, malonic, acetic and maleic acids) resulting from the 4-NP degradation. The combination of these analytical methods already led to the proposal of an oxidation/mineralization mechanism for the CWPO of 4-NP.pt_PT
dc.identifier.citationRibeiro, Rui; Silva, Adrián; Gomes, Helder; Faria, Joaquim (2015). High-performance liquid chromatography as a tool for the assessment of catalytic wet peroxide oxidation processes: in-house pre-validation of three analytical methods. U. Porto Journal of Engineering. 1, p. 50-66pt_PT
dc.identifier.urihttp://hdl.handle.net/10198/12348
dc.language.isoengpt_PT
dc.peerreviewedyespt_PT
dc.subjectHigh-performance liquid chromatography (HPLC)pt_PT
dc.subjectCatalytic wet peroxide oxidation (CWPO)pt_PT
dc.subject4-Nitrophenolpt_PT
dc.titleHigh-performance liquid chromatography as a tool to evaluate the performance of the catalytic wet peroxide oxidation of 4-nitrophenol: pre-validation of analytical methodspt_PT
dc.typejournal article
dspace.entity.typePublication
oaire.citation.conferencePlacePortopt_PT
oaire.citation.endPage66pt_PT
oaire.citation.startPage50pt_PT
oaire.citation.titleU. Porto Journal of Engineeringpt_PT
oaire.citation.volume1pt_PT
person.familyNameGomes
person.givenNameHelder
person.identifier.ciencia-id6218-1E19-13EE
person.identifier.orcid0000-0001-6898-2408
rcaap.rightsrestrictedAccesspt_PT
rcaap.typearticlept_PT
relation.isAuthorOfPublication0eb96337-224a-4339-9918-334436fbbb99
relation.isAuthorOfPublication.latestForDiscovery0eb96337-224a-4339-9918-334436fbbb99

Ficheiros

Principais
A mostrar 1 - 1 de 1
Miniatura indisponível
Nome:
U.Porto Journal of Engineering, 1 (2015) 50-66.pdf
Tamanho:
464.92 KB
Formato:
Adobe Portable Document Format
Licença
A mostrar 1 - 1 de 1
Miniatura indisponível
Nome:
license.txt
Tamanho:
1.75 KB
Formato:
Item-specific license agreed upon to submission
Descrição: