Publication
Conception, development and validation of a software interface to assess human’s horizontal intra-cyclic velocity with a mechanical speedo-meter
dc.contributor.author | Barbosa, Tiago M. | |
dc.contributor.author | Costa, M.J. | |
dc.contributor.author | Morais, J.E. | |
dc.contributor.author | Jesus, Sérgio | |
dc.contributor.author | Marques, Mário C. | |
dc.contributor.author | Batista, José | |
dc.contributor.author | Gonçalves, José | |
dc.date.accessioned | 2011-09-12T14:34:15Z | |
dc.date.available | 2011-09-12T14:34:15Z | |
dc.date.issued | 2011 | |
dc.description.abstract | The aim of this paper was to: (i) calibrate the Swimsportec® speedo-meter; (ii) concept and develop a software’s interface for it (iii) validate the full system. Calibration was done with an industrial robot. The robot arm’s performed a linear and horizontal path at a uniform movement for a set of wide range of velocities (0.5<v<4.5m/s) being the tension data acquire for each specific linear velocity. Software’s interface was developed in LabVIEW® to acquire, display and process pair wises velocity-time data on-line during the subject’s locomotion bout. To transfer data from speedo-meter to the software an acquisition card is used. System validation was done for a set of land-based human locomotion techniques (from slow walk to maximal running) in four subjects and comparing it with a Doppler radar gun. Linear regression models between speedo-meter system and radar gun where very high for both the coefficient of variation of the subject’s velocity within the full gait cycle and his maximal velocity. The 95% of interval confidence agreement limits were very close together in both variables. More than 80% of the Bland- Altman plots were with the 1.96 standard-deviation criteria used on regular basis as rule thumb for techniques validation. | por |
dc.identifier.citation | Barbosa, Tiago M.; Costa, Mário; Morais, Jorge; Jesus, Sérgio; Marques, Mário; Batista, José; Gonçalves, José (2011). Conception, development and validation of a software interface to assess human’s horizontal intra-cyclic velocity with a mechanical speedo-meter. In Congress of the International Society oh Biomechanics.Université Libre de Bruxelles & Université D´Europe. Brussels. | por |
dc.identifier.uri | http://hdl.handle.net/10198/6128 | |
dc.language.iso | eng | por |
dc.subject | Instrument | por |
dc.subject | Locomotion | por |
dc.title | Conception, development and validation of a software interface to assess human’s horizontal intra-cyclic velocity with a mechanical speedo-meter | por |
dc.type | conference object | |
dspace.entity.type | Publication | |
oaire.citation.conferencePlace | Université Libre de Bruxelles & Université D´Europe. Brussels. | por |
oaire.citation.title | Congress of the International Society oh Biomechanics. | por |
person.familyName | Barbosa | |
person.familyName | Morais | |
person.familyName | Batista | |
person.familyName | Gonçalves | |
person.givenName | Tiago M. | |
person.givenName | J. E. | |
person.givenName | José | |
person.givenName | José | |
person.identifier | 867179 | |
person.identifier | R-000-7ZW | |
person.identifier.ciencia-id | 8B11-BDC4-F6FF | |
person.identifier.ciencia-id | AA12-BF58-EE60 | |
person.identifier.ciencia-id | 8112-DCE2-D025 | |
person.identifier.orcid | 0000-0001-7071-2116 | |
person.identifier.orcid | 0000-0002-6885-0648 | |
person.identifier.orcid | 0000-0003-0344-3336 | |
person.identifier.orcid | 0000-0002-5499-1730 | |
person.identifier.rid | B-8547-2018 | |
person.identifier.scopus-author-id | 10044856400 | |
person.identifier.scopus-author-id | 20435350700 | |
person.identifier.scopus-author-id | 48361230200 | |
rcaap.rights | openAccess | por |
rcaap.type | conferenceObject | por |
relation.isAuthorOfPublication | 941a6f14-cfba-458a-a2e3-0cbd1846cd42 | |
relation.isAuthorOfPublication | 80b13e62-254d-4d46-ad90-8b509ab523a8 | |
relation.isAuthorOfPublication | 9e14e4a0-3d7a-42e8-8021-52e336b12de6 | |
relation.isAuthorOfPublication | 6a3b0b39-7fe9-4450-94f4-ced3941947da | |
relation.isAuthorOfPublication.latestForDiscovery | 6a3b0b39-7fe9-4450-94f4-ced3941947da |
Files
Original bundle
1 - 1 of 1
No Thumbnail Available
- Name:
- speedo-meter - expanded abstract-1.pdf
- Size:
- 107.18 KB
- Format:
- Adobe Portable Document Format
License bundle
1 - 1 of 1
No Thumbnail Available
- Name:
- license.txt
- Size:
- 1.71 KB
- Format:
- Item-specific license agreed upon to submission
- Description: