High-performance liquid chromatography as a tool to evaluate the performance of the catalytic wet peroxide oxidation of 4-nitrophenol: pre-validation of analytical methods

Utilize este identificador para referenciar este registo.
Nome:Descrição:Tamanho:Formato: 
U.Porto Journal of Engineering, 1 (2015) 50-66.pdf464.92 KBAdobe PDF Ver/Abrir

Orientador(es)

Resumo(s)

Nitrophenol (4-NP) in aqueous solutions with concentration in the range 0.0495-118.8 mg L-1 was developed and validated under the typical criteria for in-house pre-validations. Accordingly, linearity was demonstrated through the Fisher’s exact probability test. Accuracy and precision were then assessed in three concentration levels over the linear range. The reproducibility of the catalytic wet peroxide oxidation (CWPO) of 4-NP was evaluated together with the analytical error of 4-NP determination in independent experiments. In this case, the sum of both contributions to error never reached 3%. Two multi-component HPLC methods were also developed for the determination of possible aromatic intermediates (hydroquinone, 1,4-benzoquinone, catechol, 4-nitrocatechol and phenol) and carboxylic acids (oxalic, formic, malic, malonic, acetic and maleic acids) resulting from the 4-NP degradation. The combination of these analytical methods already led to the proposal of an oxidation/mineralization mechanism for the CWPO of 4-NP.

Descrição

Palavras-chave

High-performance liquid chromatography (HPLC) Catalytic wet peroxide oxidation (CWPO) 4-Nitrophenol

Contexto Educativo

Citação

Ribeiro, Rui; Silva, Adrián; Gomes, Helder; Faria, Joaquim (2015). High-performance liquid chromatography as a tool for the assessment of catalytic wet peroxide oxidation processes: in-house pre-validation of three analytical methods. U. Porto Journal of Engineering. 1, p. 50-66

Projetos de investigação

Unidades organizacionais

Fascículo